Sunday, April 5, 2020

MCA Terminal configuration software for MCA ver.3

https://rhelectronics.net/store/mca-terminal.html

MCA Terminal is a calibration terminal software for MCA ver.3 Multi Channel Analyzer. The software will allow you to perform channel to energy calibration of the unit, visually estimate calibration function linearity, download and view CSV data files from the unit memory, make quick erase of the entire data storage, export spectrum screenshots and send any custom command to the unit.
Current version of the software is the beta version, you are welcome to send me an email if you see any bug. The software require Microsoft .NET Framework 4.5.2 installed on your computer.

Friday, March 27, 2020

10% Discount Coupon Code: stayhome

10% Discount Coupon Code: stayhome
For purchase on rhelectronics.net website, valid by 30 April 2020.
Apply in your shopping cart before checkout.

Saturday, June 8, 2019

Gamma Spectrometer MCA with SiPM silicon photomultiplier & CsI(Tl) 6x6x50mm detector


Detector: CsI(Tl) 6x6x50mm + MicroFC 60035 SiPM 
Temperature compensated for range: 5.0C - 50.0C 
MCA ver.3.00 with internal memory and USB terminal








Tuesday, October 23, 2018

Scintillation Mode for Multi-Channel Analyzer kit


SCINTILLATION mode for searching with NaI(Tl) / CsI probe.

In firmware 1.03a we have added scintillation mode for the MCA module. The mode allows you
to use your NaI(Tl) probe for searching of low contaminated areas. First time you activate the
scintillation mode the software will run 5 minutes calibration to calculate average and standard
deviation for your probe.

Menu Functions related to Scintillation mode:

SCINT MODE – activate scintillation mode, restart applied by auto.
BEEP FREQ.HZ – set for sounder clicker frequency.
DEALAY N-pulses – the module will click on every N particle detected.
CONFIDENCE SIGMA – set 2 for fast searching or 3 for more accurate search but slower
response of alert.
CALIBRATE SD – will trigger 5 minute scintillation re-calibration on the next restart.
AVERAGING SEC – 2 or 3 seconds to average CPS results. Use 2 for faster search.

Take note, some MCA menu settings for acquisition spectrum is not active in scintillation mode!
Switch back to spectrum mode by setting SCINT MODE to OFF.